| Event Description | [Radiation Safety Officer (RSO)] for Advocate Lutheran General Hospital called Illinois Emergency Management Agency on 11/24/03 at approximately 1415 to report an event involving a Novoste Intravascular Brachytherapy [IVB] procedure. Hospital RSO stated that at approximately 1100 on 11/24/2003 during an IVB procedure with a prescribed dose of 18.4 gray, the end of the 40mm source train was not visible at the anticipated location at the end of the catheter. The sources were stuck in an apparent kink in the catheter. The source train was immediately retracted into the safe shielded position in the unit. A second attempt was then made but the sources became stuck in the same area and were again immediately retracted.
The procedure was then terminated and an analysis of the event and dose estimates were performed. An unintended area of the heart was exposed to radiation from the source train for approximately 47 seconds in the first attempt and 10 seconds in the second. The estimated radiation dose calculated to the wrong area of the heart was estimated to be approximately 5 gray. Essentially none of the prescribed dose of 18.4 gray was delivered to the intended area of the heart as the source train was retracted before reaching the intended area. The patient has been notified that there was a problem encountered during the procedure and the physician will notify the patient shortly of the particulars involved with the unintended dose delivered. The physicians do not expect any adverse medical effects from this event. Hospital RSO stated that they will carefully review this event and enhance training for this procedure. Hospital RSO added that he will notify Novoste regarding this event and submit the required written report within 15 days. Illinois Event Report ID: IL030078, License Number: IL -01152-01 Source Information: NOVOSTE Model #A1767, Serial # 91834, Radionuclide: Sr 90, Activity: .0484 Curies |
| Where | |
|---|---|
| Illinois Emergency Mgmt. Agency Illinois (NRC Region 3) | |
| Organization: | Illinois Emergency Mgmt. Agency |
| Reporting | |
| Agreement State | |
| Time - Person (Reporting Time:+15.28 h0.637 days <br />0.091 weeks <br />0.0209 months <br />) | |
| Opened: | Joe Klinger 11:32 Nov 25, 2003 |
| NRC Officer: | Gerry Waig |
| Last Updated: | Nov 25, 2003 |
| 40352 - NRC Website | |
Illinois Emergency Management Agency with Agreement State | |||
WEEKMONTHYEARENS 565742023-03-02T05:00:0002 March 2023 05:00:00 Agreement State Lost/Abandoned Source ENS 552512021-05-11T05:00:00011 May 2021 05:00:00 Agreement State Equipment Failed in Shielded Position ENS 549062020-09-22T05:00:00022 September 2020 05:00:00 Agreement State | Agreement State Report - Medical Event ENS 549052020-09-14T05:00:00014 September 2020 05:00:00 Agreement State | Agreement State Report - Loss of Yellow Ii Package Containing Approximately 61 Mci of Liquid Tl-201 ENS 548042020-07-28T05:00:00028 July 2020 05:00:00 Agreement State Agreement State Report- Patient Underdose ENS 547842020-07-14T20:30:00014 July 2020 20:30:00 Agreement State Agreement State Report - Gauge Crushed with Source Rod Extended ENS 547372020-06-04T06:00:0004 June 2020 06:00:00 Agreement State Agreement State Report - Low Level Radioactive Waste Railcar Fire ENS 547322020-05-30T12:00:00030 May 2020 12:00:00 Agreement State Agreement State Report - Damaged Gauge ENS 547022020-05-11T05:00:00011 May 2020 05:00:00 Agreement State Agreement State Report - Troxler Gauge Destroyed by Bulldozer ENS 545432020-02-21T06:00:00021 February 2020 06:00:00 Agreement State Agreement State Report - Stolen Fluorescence Analyzer Equipment ENS 545122020-02-05T06:00:0005 February 2020 06:00:00 Agreement State Agreement State Report - Lost Gauging Source ENS 545162019-02-22T06:00:00022 February 2019 06:00:00 Agreement State Agreement State Report - Reportable Equipment Failure ENS 403522003-11-24T20:15:00024 November 2003 20:15:00 Agreement State Agreement State Report - Dose to Patient Outside Intended Treatment Site 2023-03-02T05:00:00 [Table view] | |