The following report was received from the State of
New York via facsimile:
Pace Analytical Services reported that a leak test of a Varian ECD [electron capture device], model 02- 001972-00, serial no. SP472 indicated removable contamination in excess of 0.005 microcuries. Test results measured 0.0132 millicuries of contamination associated with a 15 millicurie Nickle 63 source manufactured by Isotope Products Laboratory (IPL), model NBCB14746, serial# SP472. The licensee was advised to isolate the source and not to use it. [The licensee was] requested to arrange for retesting and provide results to the NYS DOH [New York State - Department of Health]. Retesting was performed by [a Certified Health Physicist], Massachusetts License #13-3021, on July 14, 2016. The result of the second measurement was 0.0002 microcuries. Pace Analytical Services, Inc. stated that the ECO will be returned back to the manufacturer/ vendor and receipt will be sent to the NYS Department of Health.
EVENT REPORT ID NO. NYDOH - NY-16-03