The following information was received via fax:
Event Report ID No: PA090015
License No: PA-0082
Licensee: Mercy Fitzgerald Hospital
Event date and time: March 26, 2009/time unknown
Event location: Mercy Fitzgerald Hospital Lansdowne & Baily Rds., Darby, PA
Event type: Medical Event
Notifications: A letter dated April 9, 2009 was sent to DEP
Event description: Under dose for a Sr-90 eye applicator
City, State: Darby, PA
Description: PA DEP Bureau of Radiation Protection was notified in writing, dated April 9, 2009, by the Radiation Safety Officer for Mercy Fitzgerald Hospital. An ADCL calibrated Sr-90 ophthalmic applicator was used on a patient for an external eye application for the treatment of pterygium. The guard ring needed to be removed before treatment and was not, hence only delivered 5% of the prescribed dose to the patient.
The State has an inspection scheduled and will continue to keep NRC informed of the status of our investigation.
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.