The following is a summary of information provided by the
Georgia Radioactive Materials Program (the Program) via email:
The radiation safety officer (RSO) at Piedmont Athens Regional Medical Center notified the Program on August 26, 2024, that an incident occurred with Y-90 underdose. The catheter line became kinked during the procedure and the dose given was more than 20 percent below the planned dose.
The RSO will send an official written report to the Program within 15 days.
Georgia Incident Number: 86
A Medical Event may indicate potential problems in a medical facility's use of radioactive materials. It does not necessarily result in harm to the patient.
- * * UPDATE ON 9/17/2024 AT 1131 EDT FROM KAAMILYA NAJEEULLAH TO ROBERT THOMPSON * * *
The following information was provided by the Georgia Radioactive Materials Program (the Program) via email:
The Program received the official written report from the licensee radiation safety officer (RSO) on September 12, 2024. The RSO stated that one dosage was successfully delivered. The second dosage was not fully delivered due to a kink in the catheter.
Prescribed dosage: 15.68 mCi and 16.76 mCi Y-90 microspheres
Administered dosage: 5.39 mCi and 16.87 mCi Y-90 microspheres
Notified R1DO (Werkheiser),
NMSS Events Notification (email).