The following was received from the State of
Pennsylvania via email:
Notifications: The [Pennsylvania Bureau of Radiation Protection] Department's Central Office was informed of this event on March 19, 2014. This event is reportable within 24-hours per 10 CFR 30.50(b)(2).
Event Description: During a routine monthly inspection of the gauge, [the licensee] discovered that the collimator block had broken off of the shutter assembly. The collimator block was found lying in the quarter inch steel box that the gauge is housed in to protect it from the operating conditions on the battery. The cause of this event is still being investigated. Radiation surveys performed indicated no employees were exposed as a result of this event.
Manufacturer: Thermo Measure Tech
Model: 5204
Serial Number: B392
Isotope: Cs-137
Activity: 4 Ci
CAUSE OF THE EVENT: Unknown at this time.
ACTIONS: The broken shutter mechanism was removed and replaced by a licensed service provider. The shutter mechanism was tested and confirmed as operating properly. The [State of PA] Department plans a reactive inspection.
Media attention: None at this time.
PA Event Report ID No: PA140009