The following information was obtained from the state of
Mississippi via email:
Licensee reported a stuck shutter on an Ohmart/VEGA Model SH-F1 level gauge, during semi-annual checks of shutter operation. The failure was identified on 05/17/2016 by the licensee and was immediately reported to RSO. The RSO supposes stuck shutter failure was caused by age and oxidation. Licensee has contacted the manufacturer, Ohmart, for repair of shutter mechanism. Gauge remains in its normal operating position.
Isotope(s): Cs-137
Activity: 10 mCi
Source Serial No. 0441GK
Source Model No. A-2102
Mississippi Report Number: MS-16003
- * * UPDATE AT 1655 EDT ON 07/29/16 FROM H. BENJAMIN CULPEPPER TO S. SANDIN VIA EMAIL * * *
The stuck shutter has been closed and equipment involved removed and placed into a secure storage facility, on site, by an Ohmart Vega representative. There are no plans to put the unit back into service.
Mississippi closed this case as of July 29, 2016.
Notified R4DO (Pick) and
NMSS Events Notification via email.