| title = Part 21 Notification | | title = Part 21 Notification for Rosemount Transmitters | ||
| event text = The Supplier provided this notification via facsimile and is summarized below: | | event text = The Supplier provided this notification via facsimile and is summarized below: | ||
A multi-meter used in the calibration of certain pressure instruments went out of calibration between 10/10/2005 and 11/18/2005. These instruments may not conform to the published accuracy specification of +/- 0.25% of calibrated span. Rosemont Nuclear Instruments, Inc (RNII) successfully recalibrated all affected transmitters that had not yet shipped from its facility. The instruments affected by this failure are: | A multi-meter used in the calibration of certain pressure instruments went out of calibration between 10/10/2005 and 11/18/2005. These instruments may not conform to the published accuracy specification of +/- 0.25% of calibrated span. Rosemont Nuclear Instruments, Inc (RNII) successfully recalibrated all affected transmitters that had not yet shipped from its facility. The instruments affected by this failure are: |
Latest revision as of 22:19, 1 March 2018
Where | |
---|---|
Rosemount Nuclear Instruments Chanhassen, Minnesota (NRC Region 3) | |
Organization: | Rosemount Nuclear Instruments |
Reporting | |
10 CFR 21.21 | |
Time - Person (Reporting Time:+12.23 h0.51 days <br />0.0728 weeks <br />0.0168 months <br />) | |
Opened: | David Roberts 18:14 Dec 27, 2005 |
NRC Officer: | Mark Abramovitz |
Last Updated: | Dec 28, 2005 |