ENS 48757: Difference between revisions
Jump to navigation
Jump to search
StriderTol (talk | contribs) (Created page by program invented by Mark Hawes) |
StriderTol (talk | contribs) (Created page by program invented by StriderTol) |
||
Line 16: | Line 16: | ||
| event date = 02/15/2013 13:00 EST | | event date = 02/15/2013 13:00 EST | ||
| last update date = 02/15/2013 | | last update date = 02/15/2013 | ||
| title = Agreement State Report - Leaking Sealed Ni-63 Source | | title = Agreement State Report - Leaking Sealed Ni-63 Source in Gas Chromatograph | ||
| event text = The following information was received from the Commonwealth of Massachusetts via fax: | | event text = The following information was received from the Commonwealth of Massachusetts via fax: | ||
The licensee reported to the Agency (MA Radiation Control Board) that a wipe test survey of an SVAC-G module containing a Ni-63 sealed source yielded 0.0068 microcuries. The licensee gave no further details about the sealed source or the device, and stated that a written report was mailed to the Agency. The SS&D registration certificate indicates the sealed source maximum activity is 5 millicuries. The SS&D certificate indicates the source is contained inside the 'Thermo Fisher Scientific' model SVAC-G module which is contained in a 'Thermo Fisher' model EGIS Defender gas chromatograph instrument. | The licensee reported to the Agency (MA Radiation Control Board) that a wipe test survey of an SVAC-G module containing a Ni-63 sealed source yielded 0.0068 microcuries. The licensee gave no further details about the sealed source or the device, and stated that a written report was mailed to the Agency. The SS&D registration certificate indicates the sealed source maximum activity is 5 millicuries. The SS&D certificate indicates the source is contained inside the 'Thermo Fisher Scientific' model SVAC-G module which is contained in a 'Thermo Fisher' model EGIS Defender gas chromatograph instrument. |
Latest revision as of 21:01, 1 March 2018
Where | |
---|---|
Thermo Fisher Scientific, Inc. Franklin, Massachusetts (NRC Region 1) | |
License number: | 55-0447 |
Organization: | Ma Radiation Control Program |
Reporting | |
Agreement State | |
Time - Person (Reporting Time:+-4 h-0.167 days <br />-0.0238 weeks <br />-0.00548 months <br />) | |
Opened: | John Sumares 14:00 Feb 15, 2013 |
NRC Officer: | Steve Sandin |
Last Updated: | Feb 15, 2013 |
48757 - NRC Website | |