ENS 47458: Difference between revisions

From kanterella
Jump to navigation Jump to search
(Created page by program invented by Mark Hawes)
 
(Created page by program invented by StriderTol)
 
Line 16: Line 16:
| event date = 11/17/2011 15:48 EST
| event date = 11/17/2011 15:48 EST
| last update date = 11/18/2011
| last update date = 11/18/2011
| title = Agreement State Report - Patient Received Greater Than The Prescribed Dose
| title = Agreement State Report - Patient Received Greater than the Prescribed Dose
| event text = A patient had a mass surgically removed from both sides of the nose and was undergoing a topical treatment for microscopic disease using a Nucletron HDR containing 8.52 curies of Ir-192 with a Valencia Model H3 skin applicator containing a platinum filter.   
| event text = A patient had a mass surgically removed from both sides of the nose and was undergoing a topical treatment for microscopic disease using a Nucletron HDR containing 8.52 curies of Ir-192 with a Valencia Model H3 skin applicator containing a platinum filter.   
The planned administration was 7 fractions of 600 centigrays each on both the left and right side of the nose.  During the initial planning for the treatment, an error was made in calculating the dwell time for each location and this incorrect data was entered into the HDR system.  As a result of the error, the patient received 54% during each fraction (treatment) during two of the seven fractions on each side of the nose.  Prior to the third fraction, the medical team is required per procedure to check calculations and caught the error before the third fraction (treatment) was performed.   
The planned administration was 7 fractions of 600 centigrays each on both the left and right side of the nose.  During the initial planning for the treatment, an error was made in calculating the dwell time for each location and this incorrect data was entered into the HDR system.  As a result of the error, the patient received 54% during each fraction (treatment) during two of the seven fractions on each side of the nose.  Prior to the third fraction, the medical team is required per procedure to check calculations and caught the error before the third fraction (treatment) was performed.   

Latest revision as of 22:05, 1 March 2018

ENS 47458 +/-
Where
South Jersey Healthcare Regional Medical Center
Vineland, New Jersey (NRC Region 1)
License number: 450632
Organization: Nj Rad Prot And Rel Prevention Pgm
Reporting
Agreement State
Time - Person (Reporting Time:+18.95 h0.79 days <br />0.113 weeks <br />0.026 months <br />)
Opened: Bill Csaszar
15:45 Nov 18, 2011
NRC Officer: Joe O'Hara
Last Updated: Nov 18, 2011
47458 - NRC Website