ENS 49597: Difference between revisions
Jump to navigation
Jump to search
StriderTol (talk | contribs) (Created page by program invented by Mark Hawes) |
StriderTol (talk | contribs) (Created page by program invented by StriderTol) |
||
Line 16: | Line 16: | ||
| event date = 10/24/2013 EST | | event date = 10/24/2013 EST | ||
| last update date = 12/02/2013 | | last update date = 12/02/2013 | ||
| title = Agreement State Report - Equipment Failure Involving | | title = Agreement State Report - Equipment Failure Involving a Radqual Flexible Ruler | ||
| event text = The following information was received from the State of Tennessee via email: | | event text = The following information was received from the State of Tennessee via email: | ||
Tennessee's Division of Radiological Health was notified on Monday October 28, 2013 by the RSO from the University of Tennessee [UT] Knoxville, regarding the failure of a RadQual BM83-20 flexible ruler Co-57 source. In late October, a UT nuclear medicine technician put down the ruler source, and one of the rubber caps came off. This resulted in three of the metal sources inside coming out. The sources were not leaking and no radiation contamination occurred. It appeared as though the glue from the end caps had failed. The source contained 460 micro-Ci of Co-57. Manufacturer is International Isotopes Idaho, Inc., serial # BM8305710071101. The Radiation Safety Department collected the source to be disposed of in the future. | Tennessee's Division of Radiological Health was notified on Monday October 28, 2013 by the RSO from the University of Tennessee [UT] Knoxville, regarding the failure of a RadQual BM83-20 flexible ruler Co-57 source. In late October, a UT nuclear medicine technician put down the ruler source, and one of the rubber caps came off. This resulted in three of the metal sources inside coming out. The sources were not leaking and no radiation contamination occurred. It appeared as though the glue from the end caps had failed. The source contained 460 micro-Ci of Co-57. Manufacturer is International Isotopes Idaho, Inc., serial # BM8305710071101. The Radiation Safety Department collected the source to be disposed of in the future. |
Latest revision as of 20:52, 1 March 2018
Where | |
---|---|
University Of Tennessee Knoxville Knoxville, Tennessee (NRC Region 1) | |
License number: | R-47005-I18 |
Organization: | Tennessee Div Of Rad Health |
Reporting | |
Agreement State | |
Time - Person (Reporting Time:+946.97 h39.457 days <br />5.637 weeks <br />1.297 months <br />) | |
Opened: | Laura Turner 16:58 Dec 2, 2013 |
NRC Officer: | Steve Sandin |
Last Updated: | Dec 2, 2013 |
49597 - NRC Website | |