| event date = 01/04/2018 CST
| event date = 01/04/2018 CST
| last update date = 01/05/2018
| last update date = 01/05/2018
| title = Agreement State Report - Delivered Dosage Different From Prescribed
| title = Agreement State Report - Delivered Dosage Different from Prescribed
| event text = The following was received from the State of Wisconsin:
| event text = The following was received from the State of Wisconsin:
On January 4, 2018, the Department [Wisconsin Radiation Protection Section] received a telephone call and email from the licensee's medical physicist that a medical event of a prostate brachytherapy procedure was identified on January 4, 2018.  The total dose delivered differs from the prescribed dose by 20% or more.  This is a medical event as described in DHS [Department of Health Services] 157.72(1)(a)1.a.  The prescribed dose was 145 Gy; the dose delivered was 105 Gy.  The licensee uses D90 (dose delivered to 90% of the clinical target volume) < 80% for determining medical events.  Using the licensee's dose based criteria; the dose received by the prostate was 75% of the intended dose.  The post-implant computed tomography scan occurred on December 6, 2017.  The subsequent dosimetric analysis was on January 3, 2018.  The department will perform a site investigation to determine the root cause of this medical event.  
On January 4, 2018, the Department [Wisconsin Radiation Protection Section] received a telephone call and email from the licensee's medical physicist that a medical event of a prostate brachytherapy procedure was identified on January 4, 2018.  The total dose delivered differs from the prescribed dose by 20% or more.  This is a medical event as described in DHS [Department of Health Services] 157.72(1)(a)1.a.  The prescribed dose was 145 Gy; the dose delivered was 105 Gy.  The licensee uses D90 (dose delivered to 90% of the clinical target volume) < 80% for determining medical events.  Using the licensee's dose based criteria; the dose received by the prostate was 75% of the intended dose.  The post-implant computed tomography scan occurred on December 6, 2017.  The subsequent dosimetric analysis was on January 3, 2018.  The department will perform a site investigation to determine the root cause of this medical event.  

Latest revision as of 20:41, 1 March 2018

ENS 53148 +/-
Where
Aspirus-Wausau Hospital
Wausau, Illinois (NRC Region 3)
License number: 073-1342-01
Organization: Wisconsin Radiation Protection
Reporting
Agreement State
Time - Person (Reporting Time:+26.5 h1.104 days <br />0.158 weeks <br />0.0363 months <br />)
Opened: Krista Kuhlman
08:30 Jan 5, 2018
NRC Officer: Andrew Waugh
Last Updated: Jan 5, 2018
53148 - NRC Website