Line 3: Line 3:
| facility = Radamerica
| facility = Radamerica
| Organization = Maryland Dept Of The Environment
| Organization = Maryland Dept Of The Environment
| license number = Md0505103
| license number = MD0505103
| region = 1
| region = 1
| state = Maryland
| state = Maryland
Line 16: Line 16:
| event date = 12/15/2015 EST
| event date = 12/15/2015 EST
| last update date = 01/15/2016
| last update date = 01/15/2016
| title = Medical Event - Brachytherapy Source Not Placed In Intended Treatment Location
| title = Medical Event - Brachytherapy Source Not Placed in Intended Treatment Location
| event text = This is a preliminary notification by the State of Maryland, Radiological Health Program of a medical event. During both the first and second of four High Dose Rate brachytherapy treatments to one of the patient's lungs occurring on December 15, 2015 and December 21, 2015, the radioactive source was incorrectly placed between 2 and 3 centimeters away from the intended placement location. The radioactive source being used was Iridium 192. The activity level of the radioactive source is unknown at this time.
| event text = This is a preliminary notification by the State of Maryland, Radiological Health Program of a medical event. During both the first and second of four High Dose Rate brachytherapy treatments to one of the patient's lungs occurring on December 15, 2015 and December 21, 2015, the radioactive source was incorrectly placed between 2 and 3 centimeters away from the intended placement location. The radioactive source being used was Iridium 192. The activity level of the radioactive source is unknown at this time.
The placement error was discovered during the third treatment on January 13, 2016.  The reason for the placement error involved the use of an incorrect connection between the marker and the catheter.
The placement error was discovered during the third treatment on January 13, 2016.  The reason for the placement error involved the use of an incorrect connection between the marker and the catheter.

Latest revision as of 19:03, 24 November 2018

ENS 51651 +/-
Where
Radamerica
Baltimore, Maryland (NRC Region 1)
License number: MD0505103
Organization: Maryland Dept Of The Environment
Reporting
Agreement State
Time - Person (Reporting Time:+707.4 h29.475 days <br />4.211 weeks <br />0.969 months <br />)
Opened: Alan Jacobson
16:24 Jan 13, 2016
NRC Officer: Donald Norwood
Last Updated: Jan 15, 2016
51651 - NRC Website