Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 0.655 +/- 0.016 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Am-241 FWHM (keV)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Sample Mean Investigate Action Off Scale 0.58 0.6 0.62 0.64 0.66 0.68 0.7 0.72 0.74 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 A

m 2

4 1

F W

H M

Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 0.640 +/- 0.015 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Am-241 FWHM (keV)

F:\\GE UNC 081911.QAF Data Sample Mean Investigate Action Off Scale 0.56 0.58 0.6 0.62 0.64 0.66 0.68 0.7 0.72 5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 A

m 2

4 1

F W

H M

Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 1.364 +/- 0.025 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Cs-137 FWHM (keV)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Sample Mean Investigate Action Off Scale 1.25 1.3 1.35 1.4 1.45 1.5 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 C

s 1

3 7

F W

H M

Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 1.355 +/- 0.033 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Cs-137 FWHM (keV)

F:\\GE UNC 081911.QAF Data Sample Mean Investigate Action Off Scale 1.2 1.25 1.3 1.35 1.4 1.45 1.5 5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 C

s 1

3 7

F W

H M

Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 2.065 +/- 0.072 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Co-60 FWHM (keV)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Sample Mean Investigate Action Off Scale 1.7 1.8 1.9 2

2.1 2.2 2.3 2.4 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 C

o 6

0 F

W H

M Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 2.039 +/- 0.067 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Co-60 FWHM (keV)

F:\\GE UNC 081911.QAF Data Sample Mean Investigate Action Off Scale 1.7 1.8 1.9 2

2.1 2.2 2.3 2.4 5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 C

o 6

0 F

W H

M Measurement Date/Time

Lower/Upper Boundaries Selection Dates Parameter Description QA Filename 59.000 - 60.000 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Am-241Energy (keV)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Boundaries Off Scale 58.5 59 59.5 60 60.5 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 A

m 2

4 1

E n

e r

g y

Measurement Date/Time

Lower/Upper Boundaries Selection Dates Parameter Description QA Filename 59.000 - 60.000 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Am-241Energy (keV)

F:\\GE UNC 081911.QAF Data Boundaries Off Scale 58.5 59 59.5 60 60.5 5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 A

m 2

4 1

E n

e r

g y

Measurement Date/Time

Lower/Upper Boundaries Selection Dates Parameter Description QA Filename 661.100 - 662.100 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Cs-137 Energy (keV)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Boundaries Off Scale 660.5 661 661.5 662 662.5 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 C

s 1

3 7

E n

e r

g y

Measurement Date/Time

Lower/Upper Boundaries Selection Dates Parameter Description QA Filename 661.100 - 662.100 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Cs-137 Energy (keV)

F:\\GE UNC 081911.QAF Data Boundaries Off Scale 660.5 661 661.5 662 662.5 5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 C

s 1

3 7

E n

e r

g y

Measurement Date/Time

Lower/Upper Boundaries Selection Dates Parameter Description QA Filename 1331.480 - 1333.480 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Co-60 Energy (keV)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Boundaries Off Scale 1.3310e+003 1.3315e+003 1.3320e+003 1.3325e+003 1.3330e+003 1.3335e+003 1.3340e+003 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 C

o 6

0 E

n e

r g

y Measurement Date/Time

Lower/Upper Boundaries Selection Dates Parameter Description QA Filename 1331.480 - 1333.480 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Co-60 Energy (keV)

F:\\GE UNC 081911.QAF Data Boundaries Off Scale 1.3310e+003 1.3315e+003 1.3320e+003 1.3325e+003 1.3330e+003 1.3335e+003 1.3340e+003 5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 C

o 6

0 E

n e

r g

y Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 27.632 +/- 7.640 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Am-241 Decay corr. act. (uCi/unit)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Sample Mean Investigate Action Off Scale

-10 0

10 20 30 40 50 60 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 A

m 2

4 1

D e

c a

y c

o r

r a

c Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 1.116 +/- 0.313 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Am-241 Decay corr. act. (uCi/unit)

F:\\GE UNC 081911.QAF Data Sample Mean Investigate Action Off Scale

-0.5 0

0.5 1

1.5 2

2.5 5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 A

m 2

4 1

D e

c a

y c

o r

r a

c Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 25.849 +/- 7.175 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Cs-137 Decay corr. act. (uCi/unit)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Sample Mean Investigate Action Off Scale

-10 0

10 20 30 40 50 60 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 C

s 1

3 7

D e

c a

y c

o r

r a

c Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 0.896 +/- 0.160 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Cs-137 Decay corr. act. (uCi/unit)

F:\\GE UNC 081911.QAF Data Sample Mean Investigate Action Off Scale 0.2 0.4 0.6 0.8 1

1.2 1.4 1.6 5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 C

s 1

3 7

D e

c a

y c

o r

r a

c Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 25.676 +/- 7.111 8/01/11 12:00:00 AM - 8/01/12 12:00:00 AM Co-60 Decay corr. act. (uCi/unit)

G:\\DATA FORJASON\\GE UNC 040912.QAF Data Sample Mean Investigate Action Off Scale

-10 0

10 20 30 40 50 60 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 C

o 6

0 D

e c

a y

c o

r r

a c

t Measurement Date/Time

Sample Mean +/- Std Dev Selection Dates Parameter Description QA Filename 0.872 +/- 0.192 7/01/11 12:00:00 AM - 9/01/12 12:00:00 AM Co-60 Decay corr. act. (uCi/unit)

F:\\GE UNC 081911.QAF Data Sample Mean Investigate Action Off Scale 0

0.5 1

1.5 2

5/19/11 7/8/11 8/27/11 10/16/11 12/5/11 1/24/12 3/14/12 5/3/12 6/22/12 8/11/12 9/30/12 C

o 6

0 D

e c

a y

c o

r r

a c

t Measurement Date/Time