Calibration and Voltage Plateau ERG Certificate of Calibration Environmental Restoration Group, Inc.
8809 Washington St NE. Suite 150 Albuquerque. NM 87113 (505) 298-4224 www.ERGoffice.com Meter/Detector:
Manufacturer:
Ludlum Model Number:
2929 & 43-10-1 Serial Number:
157320 & PR 157821
~ Mechanical Check F/S Response Check
~ Geotropism c.__: Meter Zeroed
~ THR/WIN Operation
~ Reset Check
..; Audio Check
_ Battery Check (Min 4.4 VDC)
HV Check(+/- 2.5%):
.~: 500 V
~ 1000 V ~ 1500 V Cable Length:
~ 39-inch [,_; 72-inch lJ Other:
Alpha Threshold:
170 mV Barometric Pressure:
24.36 Inches Hg Source Distance: * :Contact Source Geometry:= Side
. 6 inches ~. Other: In planchet Beta Threshold:
4 mV Temperature:
76 l.>fl
- _:: Below
~ Other: In planchet Beta Window:
46mV Relative Humidity:
20 Instrument found within tolerance: ~ Yes
~ No Integrated I-Min. Count Integrated I-Min. Count "As Found" "Reading" Range/Multiplier Reference Setting a
~
a
~
x 1000 400 Kcpm 398969 398972 398969 398972 x 100 40 Kcpm 39898 39898 39898 39898 x 10 4 Kcpm 3989 3990 3989 3990 x I 400 cpm 399 399 399 399 High Pot.
Alpha Source Beta Source
Background
Voltage Setting a
f3 a
f3 a
f3 650 3.68 4251 299 9
2442 4
37 700 3.86 4507 306 11 3180 62 750 4.06 4945 579 10 4126 6
79 800 4.26 4934 930 7
5023 4
92 Comments: HV Plateau Scaler Count Time= I-min. Recommended HV = 700, Pot. setting= 3.86 Fluke multimeter serial number
- ]87490128
= Gamma Source Cs-137@ 5.2 uCi (1/4/12) sn: 4097-03
_ Other Source:
Calibration Date.S:--l(o-1<,,,
Calibration Due: ~~11,-11 Reviewed By:
Date:
ERG form ITC. 101.D T!11s ca/1bra11011 co11tor111s lo rh.: req111r.:111.:111s a11d accepwNe <\\1/tbra11011 co11d111011s of.1.\\ SJ.\\32.5.1 - 199"